ChipFind - документация

Электронный компонент: C4777-01

Скачать:  PDF   ZIP
Features
l Uses high sensitivity APD
Two types of APDs with different active areas
(
0.5 mm, 3.0 mm) are provided.
l Choice of high sensitivity type and low-light-level detection type
C4777 detects optical signals from 10 kHz to 100 MHz
pulsed light and is suitable for spatial light transmission and
rangefinder applications. C4777-01 provides a high photo-
electric sensitivity of 1.25 10
9
V/W and low NEP of 2 fW/Hz
1/2
,
making it excellent for fluorescence measurement and NOx
monitors where low-light-level detection is essential.
l Built-in temperature control circuit
An APD chip, thermoelectric cooler and thermosensor are
sealed in the same package along with the temperature
control circuit, to keep the APD chip temperature constant
and ensure stable measurements.
l Simple operation
Operates by just connecting to a DC power supply (+5 V, 15 V).
Applications
l Spatial light transmission
l Optical communication
l Rangefinder
l Fluorescence measurement
l NOx monitor sensor
l Particle counters
M O D U L E
APD module
APD module integrated with peripheral circuits
C4777 series
C4777
C4777-01
TEMPERATURE
CONTROL CIRCUIT
SIGNAL OUTPUT
POWER SUPPLY
FIRST AMPLIFIER
SIGNAL LIGHT
Si APD
THERMOSENSOR
THERMOELECTRIC COOLER
BIAS CONTROL CIRCUIT
HIGH VOLTAGE
POWER SUPPLY
KACCC0099EA
I Block diagram
I Selection guide
Parameter
C4777
C4777-01
Unit
Active area
f0.5
f3.0
mm
Photo sensitivity
-2.5 10
#
-1.25 10
'
V/W
Frequency bandwidth
10 k to 100 M
DC to 5 k
Hz
1
APD module
C4777 series
I Absolute maximum ratings
Parameter
Symbol
Condition
Value
Unit
Positive supply voltage
-
+7
V
Positive supply voltage
-
+16
V
Negative supply voltage
-
-16
V
Input power
-
10
mW
Operating temperature
Topr
10 to 40
C
Storage temperature
Tstg
-10 to +60
C
Operating/storage humidity
-
No condensation
70 % RH below
-
I Specification (Typ. Ta=25 C, Vcc=+5 V, +15 V, unless otherwise noted)
G Photoelectric section (APD)
Parameter
Symbol
Condition
C4777
C4777-01
Unit
Active area
A
f0.5
f3.0
mm
Peak sensitivity wavelength
lp
800
nm
Spectral response range
l
400 to 1000
nm
Photo sensitivity
S
l=800 nm, M=1
0.5
A/W
Gain
M
l=800 nm
100
50
times
Thermoelectric cooler
-
One-stage
Two-stage
-
Cooling temperature
-
0
C
Temperature stability of gain
-
Ta=10 to 40 C
3
%
G Signal amplification section
Parameter
Symbol
Condition
Min.
Typ.
Max.
Unit
High band
95
100
-
MHz
Cut-off frequency
Low band
fc
-3 dB
-
10
15
kHz
Noise equivalent power
NEP
l=800 nm
-
80
120
fW/Hz
1/2
Feedback resistance
Rf
-
10
-
k9
Photoelectric sensitivity
-
APD include, M=100
l=800 nm
-2.0
-2.5
-3.0
10
#
V/W
Output impedance
-
-
50
-
9
Maximum input light level
-
0.6
0.8
-
W
Minimum detection limit
-
-
0.8
1.2
nWr.m.s.
Parameter
Symbol
Condition
Min.
Typ.
Max.
Unit
High band
4
5
-
kHz
Cut-off frequency
Low band
fc
-3 dB
-
DC
-
-
Noise equivalent power
NEP
l=800 nm
-
2
4
fW/Hz
1/2
Feedback resistance
Rf
-
50
-
M9
Photoelectric sensitivity
-
APD include, M=50
l=800 nm
-1.0
-1.25
-1.5
10
'
V/W
Maximum input light level
-
8.8
9.6
-
nW
Minimum detection limit
-
-
0.2
0.4
pWr.m.s.
C4777
C4777-01
2
APD module
C4777 series
C4777-01
(Typ. Ta=25 C)
FREQUENCY (Hz)
PHOTO SENSITIVITY (V/W)
C4777
10
4
DC
1 k
10 k
100 k
1 M
10 M
100 M
1 G
10
5
10
6
10
7
10
8
10
9
10
10
I Spectral response
I Frequency response
I Response to stepped light
KAPDB0020EB
WAVELENGTH (nm)
PHOTO SENSITIVITY
(A/W)
(Typ. Ta=25 C,
=800 nm)
200
400
600
800
1000
40
20
0
50
30
10
M=100
M=50
KACCB0048EA
Ta=25 C, gain M=100, input pulse width=50 ns
X-axis: 10 ns/div., Y-axis: 50 mV/div.
Ta=25 C, gain M=50, input pulse width=1 ms
X-axis: 250 s/div., Y-axis: 500 mV/div.
KACCC0130EA
KACCC0131EA
I General ratings
Parameter
Min.
Typ.
Max.
Unit
+5 V
+4.4
+5
+5.6
+15 V
+14.4
+15
+15.6
Power supply
-15 V
-14.4
-15
-15.6
V
C4777 (+5 V)
-
+1.2
+1.5
C4777 (+15 V)
-
+0.2
+0.3
Current consumption
C4777 (-15 V)
-
-0.01
-0.02
A
C4777-01 (+5 V)
-
+0.5
+1.0
C4777-01 (+15 V)
-
+0.15
+0.30
Current consumption
C4777-01 (-15 V)
-
-0.01
-0.02
A
C4777
103 60 40
Dimensional outline
C4777-01
106.6 60 40
mm
Weight
350
g
C4777
C4777-01
3
APD module
C4777 series
HAMAMATSU PHOTONICS K.K., Solid State Division
1126-1 Ichino-cho, Hamamatsu City, 435-8558 Japan, Telephone: (81) 053-434-3311, Fax: (81) 053-434-5184, http://www.hamamatsu.com
U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P.O.Box 6910, Bridgewater, N.J. 08807-0910, U.S.A., Telephone: (1) 908-231-0960, Fax: (1) 908-231-1218
Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49) 08152-3750, Fax: (49) 08152-2658
France: Hamamatsu Photonics France S.A.R.L.: 8, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: 33-(1) 69 53 71 00, Fax: 33-(1) 69 53 71 10
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road, Welwyn Garden City, Hertfordshire AL7 1BW, United Kingdom, Telephone: (44) 1707-294888, Fax: (44) 1707-325777
North Europe: Hamamatsu Photonics Norden AB: Smidesvgen 12, SE-171 41 Solna, Sweden, Telephone: (46) 8-509-031-00, Fax: (46) 8-509-031-01
Italy: Hamamatsu Photonics Italia S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39) 02-935-81-733, Fax: (39) 02-935-81-741
Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions.
Specifications are subject to change without notice. No patent rights are granted to any of the circuits described herein. 2002 Hamamatsu Photonics K.K.
Cat. No. KACC1022E05
Dec. 2002 DN
60.0
40.0
37.0
30.0
57.0
45.0
9.5
2.0
106.6
60.0
29.1
5.2 0.2
POWER
SIGNAL
(4 ) M3 DEPTH 3
(FLANGE IS REMOVED.)
FRONT
SIDE
BASE
(2 ) M4 DEPTH 4
WEIGHT: 350 g
BACK
FLANGE IS REMOVED.
PHOTO-
SENSITIVE
SURFACE
POWER
SUPPLY
SIGNAL OUTPUT
ACTIVE AREA
3 mm
I Dimensional outline (unit: mm)
60.0
40.0
37.0
30.0
57.0
45.0
6.0
2.0
103.0
60.0
25.5
4.0 0.2
POWER
SIGNAL
ACTIVE AREA
0.5 mm
(4 ) M3 DEPTH 3
(FLANGE IS REMOVED.)
FRONT
SIDE
BASE
(2 ) M4 DEPTH 4
POWER
SUPPLY
WEIGHT: 350 g
SIGNAL OUTPUT
BACK
FLANGE IS REMOVED.
PHOTO-
SENSITIVE
SURFACE
KACCA0071EA
KACCA0080EA
When using C4777-01, do not terminate the output with a 50
load.
When using C4777, be sure to terminate the output with a 50
load.
C4777
C4777-01
4