ChipFind - документация

Электронный компонент: C4878

Скачать:  PDF   ZIP


High Performance Thermoelectric Coolers C4877, C4878 Series
Specifications
[Cooled PMT Housing]
[Power Supply]
[Components and Accessories]
GCooled PMT Housing (Including an input window) GPower Supply GLight Shield Cap
GSpare Fuse GWater Hose Clamps GConnection Cable (1.5 m) GAC Line Cable (2 m)
Features
G Thermoelectric cooling using Peltier elements
G About -30 C cooling temperature (with +20 C cooling water)
G Evacuated, double-pane window with heater for frost prevention
G Built-in electrostatic and magnetic shielding (C4877 Series)
G Water shut-off protection to guard the Peltier elements
G Stable operation due to a regulated power supply
102
The C4877 series and C4878 series are thermoelectric coolers constructed
with enhanced electrostatic and magnetic shielding (C4877 Series). This
minimizes the influence of external noise on the photomultiplier tube and
thus significantly improves photometric accuracy. These coolers offer user-
friendly functions such as easy temperature control and pilot lamp blank-
ing. The C4877 series is designed for use with 51 mm (2") or 38 mm
(1-1/2") diameter head-on photomultiplier tubes, and the C4878 series for
TACCF0161
Cooling
Heat Exchange Medium
Amount of Cooling Water
Cooling Temperature (with cooling water at +20
C)
Temperature Controllable Range (with cooling water at +20
C)
Cooling Time
Optical Window Material
Applicable PMTs (Optional)
Applicable Socket Assembly
or PMT Holder (Optional)
Weight
C4877 Series
C4878 Series
C4877 Series
C4878 Series
Parameter
Value/Description
Thermoelectric cooling using Peltier elements
Water
1 L/min to 3 L/min
Approx. -30
C
-30
C to 0 C (continuously adjustable)
Approx. 120 min
Evacuated double-pane fused silica window with heater
28 mm (1-1/8") Dia. and 38 mm (1-1/2") Dia. and 51 mm (2") Dia. Head-on
MCP-PMT (R3809U-50 Series)
E2762 Series
A
E3059-500 (R3809U-50 Series)
5.8 kg
NOTE: A
Left: C4877 Power Supply
Right: C4877 Cooled PMT Housing
Input Voltage (AC)
Power Consumption
Output Voltage
Output Current
Protection Circuit
Weight
C4877, C4878
C4877-01, C4878-01
C4877-02, C4878-02
Parameter
Value/Description
100 V
10 V (50 Hz/60 Hz)
120 V
12 V (50 Hz/60 Hz)
230 V
23 V (50 Hz/60 Hz)
270 VA
28 V
4.3 A
Functions against cooling water suspension and over current/short circuit
8.5 kg
E2762 Series
E2762-502
E2762-506
E2762-509
E2762-510
E2762-511
E2762-513
R1767
R943-02, R3310-02
R464, R585, R649
R329, R331
R316-02, R374, R2228, R5929, R6249
R375, R669
PMT
All 6 models E2762 series listed here are
SHV high voltage input and BNC current output type.
To operate C4877 and C4878, water hoses with an inner diameter of 15 mm are required.
Thermoelectric Coolers
Dimensional Outlines
(Unit: mm)
Cooling Characteristics
TACCA0172EB
TACCA0173EB
TACCB0034EA
TACCB0035EA
Cooling Effect on Dark Count
(C4877)
Spectral Transmission Characteristics
of Optical Window
TACCB0038EB
TACCB0036EA
0
1000
CHANNEL
DARK COUNT (s
-1
)
10
0
500
10
1
10
2
10
3
10
4
10
5
PMT : R943
-02
ROOM TEMPERATURE
(17400 s
-1
)
-25 C (4 s
-1
)
100
1200
400
1000
WAVELENGTH (nm)
TRANSMITTANCE (%)
0
40
100
200
600
800
20
60
80
6-M3 THREAD
S-100 O-RING
WINDOW FLANGE
180
1041.5
160
215
275
8
35 MAX.
16
30
126
140
200
302
205
140
35
30
200
8
50
+2
0
WINDOW FLANGE
PMT
EVACUATED WINDOW
HOUSING
FRONT PANEL
HOUSING
POWER SUPPLY
REAR
(C4877 Series)
FRONT
52
86
95
100
130
120
(C4878 Series)
6-M3 THREAD
S-100 O-RING
WINDOW FLANGE
130
120
61.5
PHOTOCATHODE
8
52
86
95
100
WINDOW FLANGE
PMT
EVACUATED WINDOW
HOUSING
FRONT PANEL
* C4877-02 and C4878-02 conform to the EMC directive (89/336/EEC) and the LVD (73/223/EEC) of the European Union.
0
120
40
80
TIME (min)
COOLING TEMPERATURE (C)
-40
-30
100
20
60
-20
-10
0
+10
+20
+30
COOLING WATER
AMBIENT TEMPERATURE
:
+20 C
:
+20 C
0
+30
+10
+20
COOLING WATER TEMPERATURE (C)
COOLING TEMPERATURE (C)
-50
-30
-20
-10
-40
103
104
Thermoelectric Coolers C659 Series
Specifications
[Cooled PMT Housing]
[Power Supply]
[Components and Accessories]
GCooled PMT Housing (Including a magnetic shield and input window) GPower Supply GLight Shield Cap
GConnection Cable (2 m) GProtective Circuit Cable (2 m) GAC Line Cable (2 m)
Features
G Thermoelectric cooling using Peltier elements
G Evacuated, double-pane window with heater for frost prevention
G Water shut-off protection to guard Peltier elements
G Built-in magnetic shield
The C659 series is a thermoelectric cooler designed to reduce photomultipli-
er tube dark current and to enhance the photomultiplier tube lower detection
limit. When used with a photomultiplier tube having an Ag-O-Cs photoca-
thode in DC mode, the dark current can be reduced to 1/200 of its normal
level at room temperature.
In cooling a photomultiplier tube one sometimes encounters problems such
as dewing on the incident window and leakage current on the socket or base
of the tube. But these problems are eliminated by use of an evacuated, dou-
ble pane window and custom socket assembly.
The C659-50 series is designed for 38 mm (1-1/2") and 28 mm (1-1/8") di-
ameter head-on photomultiplier tubes. The C659-70 series is specifically in-
tended for use with 28 mm (1-1/8") diameter side-on photomultiplier tubes.
Cooling
Heat Exchange Medium
Amount of Cooling Water
Cooling Temperature
A
Cooling Time
Optical Window Material
Applicable PMTs
Applicable Socket Assemblies
Weight
C659-50 Series
C659-70 Series
C659-50 Series
C659-70 Series
C659-50 Series
C659-70 Series
C659-50 Series
C659-70 Series
Parameter
Value/Description
Thermoelectric cooling using Peltier elements
Water
1 L/min to 3 L/min
Approx. -20
C
Approx. -15
C
Approx. 60 min
Evacuated double-pane fused silica window with heater
38 mm (1-1/2") Dia. and 28 mm (1-1/8") Dia. Head-on
28 mm (1-1/8") Dia. Side-on
E1135-500, -501, -502 (Optional)
A
E1135-503 (Supplied)
A
4.8 kg
2 kg
To operate C659 series, water hoses with an inner diameter of 15 mm are required.
Input Voltage (AC)
Power Consumption
Output Voltage
Output Current
Protection Circuit
Weight
Parameter
Value/Description
100 V
10 V (50 Hz/60 Hz)
115 V
12 V (50 Hz/60 Hz)
220 V
22 V (50 Hz/60 Hz)
160 VA
110 VA
6 V
3.4 V
12 A
12 A
Functions against cooling water suspension
10.7 kg
8.5 kg
NOTE:
A Cooling water: +20 C, Cooling temperature differs depending on photomultiplier tube type.
C659-50, -70
C659-51, -71
C659-52, -72
C659-50 Series
C659-70 Series
C659-50 Series
C659-70 Series
C659-50 Series
C659-70 Series
C659-50 Series
C659-70 Series
LLeft: C659-50 Series Right: C659-70 Series
NOTE:
A E1135 Series
E1135-500
E1135-501
E1135-502
E1135-503
R980, R1387
R580
R6248
28 mm Dia. Side-on
PMT
All 4 models E1135 series listed here are
SHV high voltage input and BNC current output type.
Thermoelectric Coolers
105
Dimensional Outlines
(Unit: mm)
Cooling Characteristics
(C659-50 Series)
Effect by Cooling Water Temperature
(C659-50 Series)
TACCA0135EC
TACCA0137EB
TACCB0033EA
TACCB0029EA
G Cooled PMT Housing (C659-50 Series)
G Power Supply
TACCA0136EB
G Cooled PMT Housing (C659-70 Series)
-40
-20
+20
0
COOLING TEMPERATURE (

C)
0
162
40
162
110
(FRONT)
23.5
110
COOLING WATER IN/OUT
(REAR)
(SIDE)
15
40
280
35
SOCKET
ASSEMBLY
PMT
81
EVACUATED WINDOW
15
200
41
M80 P=1
M4
SIGNAL
HV
126
46
11.5
171
32
60
M60 P=1
35
15
100
COOLING
WATER IN/OUT
SOCKET
ASSEMBLY
PMT
7
24
8
EVACUATED
WINDOW
M56 P=0.75
95
20
M4
80
290
20
260
C659-50 SERIES...130
C659-70 SERIES...125
C659-50 SERIES...150
C659-70 SERIES...130
C659-50
SERIES...120
C659-70
SERIES...100
TERMINAL BASE
-40
-20
-10
+10
+20
0
20
40
60
TIME (min)
TEMPERATURE (

C)
0
-30
INTERNAL
WALL TEMPERATURE
OF PMT HOUSING
PMT TEMPERATURE
COOLING WATER: +20
C
-40
-20
+20
0
+40
COOLING WATER TEMPERATURE (
C)
COOLING TEMPERATURE (

C)
0
+20