ChipFind - документация

Электронный компонент: C7700

Скачать:  PDF   ZIP
High Dynamic Range Streak Camera
C7700
v
The relaxation oscillation of a laser diode can be measured
with the C7700 under single shot operation.
(The data was captured with a C4880-21 readout CCD camera.)
Less than 60 ps
The C7700 is a newly developed high dynamic range
streak camera that can handle a large number of photo-
electrons. This feature enables single-shot
measurements of ultrafast phenomena with a D-range
as high as 10 000 : 1.
This system is suitable for high-precision simultaneous
measurement of high-intensity and weak intensity pulse
light.
10 000 :1
7
Readout System
(HPD-TA)
6
Readout Cameras
C7700
GP-IB board
Frame Grabber
Streak Image
Analysis
Software
Streak Image
Analysis Systems
for IBM
PC/AT
Dual-Scan Cooled CCD Camera C4880-21
Camera head
Control unit
1
Input Optics
3
Main Unit
Mount Table A7703
5
Output Optics
ORCA
2
Camera head
Control unit
Personal computer
(IBM PC/AT Compatible)
DUAL MODE COOLED CCD CAMERA
POWER
High Dynamic Range Measure
APPLICATIONS
q
Research involving laser fusion lasers, free
electron lasers and various other types of
pulsed lasers
q
Plasma light emission, radiation, laser ab-
lation, combustion and explosion
q
Picosecond transient absorption measurement
(Time dependence of absorption is shown below.)
q
LIDAR, Thomson scattering, time-of-flight
laser ranging
q
Fluorescence lifetime measurement, time-
resolved Raman spectroscopy
SYSTEM CONFIGURATION
Optical
intensity
Time
Space
Trigger signal
Slit
Lens
Incident light
Photocathode
(light
electrons)
Accelerating electrode
(where electrons are
accelerated)
Phosphor screen
(electrons
light)
Sweep circuit
Streak image
Time
Space
2
Streak Tube
4
Image Intensifier,
Image Booster
Sweep electrode
(where electrons are
swept in the direction
from top to bottom)
[Operating Principle]
The light pulse to be measured is focused onto the photocathode
of the streak tube through the slit, where the photons are con-
verted into a number of electrons proportional to the intensity of
the incident light. These electrons are accelerated and conducted
towards the phosphor screen, and a high-speed voltage which is
synchronized to the incident light is applied. The electrons are
swept at high speed from top to bottom, after which they are bom-
barded against the phosphor screen of the streak tube and con-
verted to an optical image.
When the light intensity of the streak image is very weak, an image
intensifier or an image booster amplifies the low light level streak
image.
OD = 0.1
0
1
2
3
4
5
Time (ns)
S
1
- S
n
absorption
=
482 to 489 nm
T
1
- T
n
absorption
=
557 to 581 nm
v
The 266 nm excitation (25 ps, 0.2 mJ, 2 mm focused,
single shot) of Chrysene in THF (0.5 mmol/l)
D-range of 10 000:1 of ultrafast phenomena under single-shot operation
FEATURES
q
High dynamic range of 10 000:1*
q
Temporal resolution of 5 ps*
q
Effective photocathode size: 17 mm
q
Simultaneous measurement of light intensity
on temporal and spatial (wavelength) axis
q
IEEE 488 (GP-IB) control
*[About a Dynamic Range]
The maximum dynamic range of the streak camera essence is set to
1000:1 by the measurement condition of temporal resolution 5ps, and
is set to 10 000:1 by the measurement condition of temporal-resolution
100ps. With a light intensity, the image intensifier tube, such as an
image intensifier or an image booster, is combined with a streak
camera. In this case, the image intensifier tube may restrict the
dynamic range of the measurement system. Furthermore, a read-out
camera may also restrict the dynamic range of the measurement
system.
Image Intensifier Tube
C4880-21
ORCA
2
Readout Camera
Image Intensifier
Image Intensifier
Image Booster
Image Booster
Temporal Resolution
5 ps
100 ps
1000:1
1000:1
-
-
2000:1
2000:1
10 000:1
3000:1
ment Possible up to 10 000:1
3
SPECIFICATIONS
1
Input Optics
Type
No.
Overall
Length
Slit Width
Reading
Precision
Image
Multiplica-
tion Ratio
Spectral
Transmission
Effective
F Value
Slit
Width
A1974
400 nm to 900 nm
1.2
1 : 1
0 to 5 mm
5
m
159 mm
A1976-04
200 nm to 1060 nm
3.5
1 : 1
98.2 mm
2
Streak Tube
100
80
60
40
20
0
200
400
600
A1974
A1976-04
800
Wavelength (nm)
Spectral transmittance of input optics
Transmittance (%)
1000
1200
1400
1600
4
Image Intensifier (I.I.), Image Booster (I.B.)
6
Readout Cameras
5
Output Optics
Photocathode
V3346U-04
(I.I.)
S-20
25 mm
P-43
3000
V3346U-24
(I.B.)
15 (typ.)
Effective
Photocathode Size
Phosphor
Screen
Luminous
Gain
A7702-02
Magnification
Readeout Camera
Type No.
Imaging Device
C4880-21
Full frame transfer CCD
512(H)
512 (V)
24
m
24
m
ORCA
2
Progressive-scan interline CCD
1280(H)
1024(V)
6.7
m
6.7
m
Effective No. of Pixels
Pixel Size
Type No.
7
Readout System
q
HPD-TA
C7700-01
C7700-11
Type
No.
Effectiv e
Photocathode
Size
N8059
200 nm to 850 nm
7 mm
17 mm
N8059-02 300 nm to 1060 nm
18 lp/mm or
more
centered on
photocathode
Note) read-
out camera
ORCA
2
Phosphor screen
characteristic P-43
Fiber-optic output
Effective Phos-
phor screen size:
25 mm
Phosphor
Screen
Spatial
Resolution
Spectral
Response
Characteristic
Streak
Tube
10
5
10
4
10
3
10
2
10
1
200
400
600
Wavelength (nm)
Spectral response characteristics
Radiant sensitivity (
A/W)
800
1000
1200
C7700-11
C7700-01
3
Main Unit
q
Streak Unit
Temporal resolution:
better than 5 ps (at the fastest sweep
range)
Sweep time/full screen
0.5, 1, 2, 5, 10, 20, 50, 100, 200, 500 ns
1, 2, 5, 10, 20, 50, 100, 200, 500
s, 1 ms
Trigger jitter:
less than
20 ps (at the fastest
sweep range)
Trigger delay:
about 25 ns (at the fastest sweep range)
Maximum sweep repetition frequency
.......................................... 1 kHz (at the fastest sweep range)
Operating mode ................................................. Focus/Streak
Trigger input
Maximum input voltage ...........................................
5 V/50
Trigger level ....................................................
4 V Adjustable
Gating mode ............................. NORMAL, Gate, Open Fixed
Gating method ...................................... Photocathode Gating
Gate trigger input ......................................... 3.5 Vto 5.0 V/50
Gate
repetition
frequency ............................................ 100 Hz
Gate delay time
..............................................................
1
s
Line voltage
Power consumption ................................... Approx. 180 V
A
Operating temperature ................................. +10 C to +30 C
Operating humidity
Less than 70 % (with no condensation)
UTILITY
The HPD-TA (Temporal Analayzer) is a high-perfomance digital
data acquisition and control system specifically designed to read-
out images from the phosphor screen of the HAMAMATSU
streak camera. It enables precise, quantitative acquisition and
preanalysis of two dimensional streak data that includes photon
counting plus a full range of data correction and calibration
possibilities. The HPD-TA allows the remote control of the
C7700 via GPIB interface. The entire system is controlled
through a powerful and user-friendly application software that
runs on a Microsoft Windows platform
C4880-21
Effective F-number
1 : 0.7
F = 2.0
Type No.
VA100 V to 120/220 V to 240 C
.........................
.....
A2098
ORCA
2 : 1
F = 2.5
2
OPTIONAL
5
IBM is a registered trademark of IBM Co.
5
Windows is a trademark of Microsoft Corporation in the U.S.A.
5
is a registered trademark of Apple Computer, Inc.
Cat. No. SSCS1049E07
APR/2002 HPK
Created in Japan (PDF)
Delay Unit C1097-01
This passive delay unit provides convenient timing adjustment with
zero jitter. Optional remote GPIB control (C1097-04).
Variable Delay Range
0 to 31.96 ns
Delay Setting Range
30, 60, 120, 250, 500 ps, 1, 2 4, 8, 16 ns
Minimum Delay Time
Approx. 12 ns
Maximum Input Voltage 30 V
Power Supply
Dimensional Outline
215 (W)
350 (D)
102 (H) mm/3.4 kg
q
q
PIN Photodiode C1083-01
Converts low-repetition light pulses to an electronic trigger for streak
sweep.
Spectral response
400 to 1100 nm
Rise time
0.8 ns
Power supply
+22.5 V (battery)
Dimensional Outline
Head: 100 (W)
160 to 235 (H)
50 (D) mm/400 g
Power supply unit: 100 (W)
83(H)
100 (D) mm/400 g
DIMENSIONAL OUTLINES
(Unit: mm)
q
Main Unit (approx. 24.0 kg)
285
1
(C-mount)
160.5
3
106.6
1
390
1
40.5
1
250
1
160
0.5
85
0.5
310
1
HIGH DYNAMIC RANGE STREAK CAMERA
C7700
A1976-04
A2098
5
Product and software package names noted in this documentation are trademarks or registered trademarks of their respective manufacturers.
q
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office.
q
Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions.
Specifications and external appearance are subject to change without notice..
2002 Hamamatsu Photonics K.K.
Homepage Address http://www.hamamatsu.com
HAMAMATSU PHOTONICS K.K., Systems Division
812 Joko-cho, Hamamatsu City, 431-3196, Japan, Telephone: (81)53-431-0124, Fax: (81)53-435-1574, E-mail:export@sys.hpk.co.jp
U.S.A. and Canada: Hamamatsu Photonic Systems: 360 Foothill Road, Bridgewater, N.J. 08807-0910, U.S.A., Telephone: (1)908-231-1116, Fax: (1)908-231-0852, E-mail: usa@hamamatsu.com
Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658, E-mail: info@hamamatsu.de
,
France: Hamamatsu Photonics France S.A.R.L.: 8, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10, E-mail: infos@hamamatsu.fr
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road, Welwyn Garden City, Hertfordshire, AL7 1BW, U.K., Telephone: (44) 1707-294888, Fax: (44) 1707-325777, E-mail: info@hamamatsu.co.uk
North Europe: Hamamatsu Photonics Norden AB: Smidesvgen 12, SE-171-41 Solna, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01, E-mail: info@hamamatsu.se
-
Italy: Hamamatsu Photonics Italia S.R.L.: Strada della Mois, 1/E 20020 Arese (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741, E-mail: info@hamamatsu.it
ISO 9001
Certificate: 09 105 79045
AC100 V to 240 V